Emerson NGA2000 Manuale Utente
6 - 26
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Main Menu — Technical level configuration
↓↓
Diagnostic menus
↓↓
Control module diagnostics
↓↓
In the menu "Control Unit Diagnostics" you can find software error messages of the
control module. If such messages are available, you can act in the following way:
control module. If such messages are available, you can act in the following way:
1) Make a note of the error messages.
2) Reset the message:
•
Press the ENTER key or the
→
→
-key in the line "Edit to reset".
•
Set up "Reset" with the
↑↑
-key or the
↓↓
-key and confirm it with the ENTER key:
The error messages will disappear if the reason is not existing any longer.
•
If the error messages keep displayed:
3) System reset:
•
Change to the menu "System Reset" with the F3 key (REBOOT).
•
Press the ENTER key or the
→
→
6-14).
4) Check the error messages:
•
Go again to the menu "Control Unit Diagnostics".
•
If the error messages are still existing, contact your customer service, please.
•
If the error messages are disappeared, you should set up the report status again:
Set up "Report" in the line "Edit to reset".
Set up "Report" in the line "Edit to reset".
HOME
ESCAPE
REBOOT
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
Control Unit Diagnostics
Software error code (1 = no error):
1
And:
And:
And:
And:
And:
And:
And:
And:
And:
And:
Edit to reset:
Report
Last software error message
And:
6.3.1 Diagnostic Menus
Control Module Diagnostics