Emerson NGA2000 Manuale Utente

Pagina di 178
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
6 - 39
Main Menu — Technical level configuration
↓↓
Diagnostic menus
↓↓
Analyzer module diagnostics
↓↓
Trend display control
↓↓
The analyzer stores 24 hours of 15 minute averages. In the menu "Trend display control"
you can make these values available to a PC for two different variables.
Note: The trend display control is not available at the moment !
Set up parameters:
♦ 
Select any line of variables with the 
↓↓
 -key or the 
↑↑
 -key.
♦ 
Select the variable with the ENTER key or the 
 -key.
♦ 
Select the parameter you want with the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
Lines of variables "First/Second displayed variable":
In these two lines you can set up one of the following parameters:
♦ 
Sample pressure
♦ 
Sample flow
♦ 
Concentration
♦ 
Noise level (of concentration)
♦ 
Raw signal
HOME
ESCAPE
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
Trend display control
First displayed variable:
Concentration
Timebase:
24  Hrs
Drop out to measuring mode:
Disabled
Second displayed variable:
Sample flow
6.3.2.7 Diagnostic Menus
Trend display control