Emerson NGA2000 Manuale Utente

Pagina di 178
System Calibration
90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98
NGA 2000
Supplement - 5
It is also possible to configure the DIO-outputs via LON-Variables (i.e. SLTA-adapter or
AK-protocol).
Therefore is the LON-variable "DIOOUT_ENTRYSIG" which corresponds to
DIOOUTSIGC and evaluates enum-values of ST1NAME (enum-values 0..19) or
ST2NAME (enum-values 20..39).
For digital output we have to setup variables in following order.
1. DIOSLOTIDC
2. DIOOUTNUMC
3. DIOOUTSRCC
4. DIOOUT_ENTRYSIG
SYS: VALVE-1 
= 20
(enum-value in ST2NAME)
 
SYS: VALVE-2
= 21
 
.
    .
 
.
    .
    
.
    .
 
SYS:VALVE-16
= 35
 
For each needed system valve we have to repeat these appropriate selections.
It is also possible to distribute the system valves onto different output modules.