Emerson NGA2000 Manuale Utente
5 - 14
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Main Menu — Expert controls and set up
↓↓
Analyzer module set up
↓↓
Gas measurement parameters
↓↓
Response time/delay parameters
↓↓
In the menu "Response time/delay parameters" you can set up the response time (t
90
time) of the analyzer for each range and the output delay of the signal.
Set up parameters:
♦
Select any line of variables with the
↓↓
-key or the
↑↑
-key.
♦
Select the variable with the ENTER key or the
→
→
-key.
♦
Select any digit with the
←
←
-key or the
→
→
-key and adjust a new value with
the
↑↑
-key or the
↓↓
-key
resp. select the whole parameter with the
↑↑
-key or the
↓↓
-key.
♦
Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
cancel and go back to the last value with the F2 key.
Lines of variables "Range 1/2/3/4 t90 time":
The t
90
time is defined as the time that has to run until the analyzer displays 90% of the
current concentration after a jump in the concentration level.
Options: 0 to 30 seconds for every range of a channel.
Options: 0 to 30 seconds for every range of a channel.
HOME
ESCAPE
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
Response time/delay parameters
Range 1 t90 time:
3.0 s
Range 3 t90 time:
3.0 s
Range 4 t90 time:
3.0 s
LON update rate:
10 per sec
Output delay time:
0.0 s
Range 2 t90 time:
3.0 s
5.1.5 Analyzer Module Set up
Response Time/Delay Parameters