Emerson NGA2000 Manuale Utente

Pagina di 178
5 - 14
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Main Menu — Expert controls and set up
↓↓
Analyzer module set up
↓↓
Gas measurement parameters
↓↓
Response time/delay parameters
↓↓
In the menu "Response time/delay parameters" you can set up the response time (t
90
time) of the analyzer for each range and the output delay of the signal.
Set up parameters:
♦ 
Select any line of variables with the 
↓↓
 -key or the 
↑↑
 -key.
♦ 
Select the variable with the ENTER key or the 
 -key.
♦ 
Select any digit with the 
 -key or the 
 -key and adjust a new value with
the 
↑↑
 -key or the 
↓↓
 -key
resp. select the whole parameter with the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
Lines of variables "Range 1/2/3/4 t90 time":
The t
90
 time is defined as the time that has to run until the analyzer displays 90% of the
current concentration after a jump in the concentration level.
Options: 0 to 30 seconds for every range of a channel.
HOME
ESCAPE
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
Response time/delay parameters
Range 1 t90 time:
3.0 s
Range 3 t90 time:
3.0 s
Range 4 t90 time:
3.0 s
LON update rate:
10 per sec
Output delay time:
0.0 s
Range 2 t90 time:
3.0 s
5.1.5 Analyzer Module Set up
Response Time/Delay Parameters