Emerson NGA2000 Manuale Utente

Pagina di 178
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 43
♦ 
Go to the line you want with the 
↑↑
 -key or the 
↓↓
 -key. If there are more than eight source
modules available you have to press the F3 key to go to another menu page.
♦ 
Select the tag of the module with the ENTER key or the 
 -key:
The display will go back to the menu "DIO-MODULE OUTPUTS" automatically.
The tag of the module selected will appear in the line "Signal comes from".
3) Select any signal:
♦ 
Change to the line "Choose signal..." with the 
↓↓
 -key.
♦ 
Press the ENTER key or the 
 -key to change to the submenu "Signals":
♦ 
Go to the line you want with the 
↓↓
 -key or the 
↑↑
 -key. With the F3 key you can go to
other menu pages containing further available signals:
• 
RANGE OVERFLOW
• 
RANGE UNDERFLOW
• 
FLOW TO LOW
• 
FLOW TO HIGH
• 
RANGE 1, 2, 3, 4
• 
OZONATOR OFF
• 
NOx MODE
• 
NO MODE
• 
and further signals, if need be:
All signals of the control module and of the analyzer modules joint to the platform
or MLT analyzer (e.g. CLD, FID, MLT AM) are available in the menu "SIGNALS".
The signals shown above will appear by using a CLD module as source module.
♦ 
Select the signal with the ENTER key or the 
 -key:
The display will go back to the menu "DIO-MODULE OUTPUTS" automatically. The
signal selected will appear in the line "Signal name".
HOME
ESCAPE
MORE
BACK
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
Signals
NORMAL
MAINTENANCE REQUEST
FAILURE
CAL. IN PROGRESS
ZERO IN PROGRESS
ZERO FAILED
SPAN FAILED
SPAN IN PROGRESS