Notice for this monitor
This product is suited to display medical images of such modalities as MRI and CT. It does not support the display of 
mammography images for diagnosis.
This product has been adjusted specifically for use in the region to which it was originally shipped. If the product is 
used outside the region, it may not operate as specified in the specifications.
This product may not be covered by warranty for uses other than those described in this manual.
The specifications noted in this manual are only applicable for power cords and signal cables specified by us.
Use optional products manufactured or specified by us with this product.
As it takes about 30 minutes for the performance of electrical parts to stabilize, adjust the monitor 30 minutes or more 
after the monitor power has been turned on.
In order to suppress the luminosity change by long-term use and to maintain the stable luminosity, use of a monitor in 
lower brightness is recommended.
When the screen image is changed after displaying the same image for extended periods of time, an afterimage may 
appear. Use the screen saver or power save function to avoid displaying the same image for extended periods of time.
Periodic cleaning is recommended to keep the monitor looking new and to prolong its operation lifetime. (Refer to 
” on the next page.)
The screen may have defective pixels. These pixels may appear as slightly light or dark area on the screen. This is due 
to the characteristics of the panel itself, and not the product.
The backlight of the LCD panel has a fixed life span. When the screen becomes dark or begins to flicker, please 
contact your dealer.
Do not press on the panel or edge of the frame strongly, as this may result in the display malfunction, such as the 
interference patterns, etc. If pressure is continually applied to the LCD panel, it may deteriorate or damage your LCD 
panel. (If the pressure marks remain on the LCD panel, leave the monitor with a white or black screen. The symptom 
may disappear.)
Do not scratch or press on the panel with any sharp objects, such as a pencil or pen as this may result in damage to the 
panel. Do not attempt to brush with tissues as this may scratch the LCD panel.
When the monitor is cold and brought into a room or the room temperature goes up quickly, dew condensation 
may occur inside and outside the monitor. In that case, do not turn the monitor on and wait until dew condensation 
disappears, otherwise it may cause some damages to the monitor.