9.3
 
  REAL
 
TIME
 
CLOCK
 
SETUP 
The  Real  Time  Clock  (RTC)  is  used  only  for  the  RS-232  output.    The  Date  and  Time 
can be set as required.  The scale will keep the clock running even when the power 
is off. 
Setting up the clock  
 
Press and hold the [CE] key for 4 seconds, release when the date and time is 
displayed.  The initial displays show the current date and time set.   
“ rtC       “  “11,14,06”   “16,41,35” 
 
Press the [CE] key to change the date and time.  The display will show the 
current time in the format, “H-m-S”. 
“timE”  “H-m-S”   “142929”  or if no time has been set show “000000”   
 
Enter the time using the numeric keys using a 24 hour clock format, 3:41PM is 
“154100”. 
“timE”  “H-m-S”   “154100”  new time overwrites the old 
 
Press  the  [Tare]  key  to  accept  the  time.    The  display  will  show  the  current 
date format.   
“dAtE”  “y-m-d” 
 
Press the [U.Wt./Unit] key to change the date format. Available formats are: 
“Y-m-d”   year, month, day 
“m-d-Y”   month, day, year 
“d-m-Y”   day, month, year 
 
Press the [Tare] key to accept the chosen format and then enter the date in 
this format.   
“dAtE”    “m-d-y”  “051209”    note  current  date  is  in  current  format,  will 
show 000000 if no date has been set. 
 
Press the [Tare] key to accept the date. 
An  error  code  will  be  shown  if  the  time  (Err  1)  or  the  date  (Err  2)  is  not  the 
permissible values.   For example, 34th day of a month is an invalid entry.