Agilent Technologies 664xA Manuale Utente

Pagina di 70
40     Language Dictionary
OUTP:REL
 
 Do not install or program the Agilent Relay Accessories if the power supply maximum output current
rating (see Table 3-1) exceeds the contact ratings of the relay.
 
 
 This command is valid only if the power supply is configured for the optional relay connector.  Programming ON closes the
relay contacts; programming OFF opens them.  The relay is controlled independently of the output state.  If the power
supply is supplying power to a load, that power will appear at the relay contacts during switching.  If the power supply is not
configured for the relay option, sending either relay command generates an error.
 
 
Command Syntax
OUTPut:RELay[:STATe] <bool>
 
Parameters
0 | 1 | OFF | ON
 
*RST Value
0
 
Examples
OUTP:REL 1     OUTP:REL OFF
 
Query Syntax
OUTPput:RELay?
 
Returned Parameters
0 |1
 
Related Commands
OUTP[:STAT] *RCL *SAV
 
OUTP:REL:POL
 
 Do not install or program the Agilent Relay Accessories if the power supply maximum output current
rating (see Table 3-1) exceeds the contact ratings of the relay.
 
 
 This command is valid only if the power supply is configured for the optional relay connector.  Programming NORMal
causes the relay output polarity to be the same as the power supply output.  Programming REVerse causes the relay output
polarity to be opposite to that of the power supply output.  If OUTP[:STAT] = ON when either relay command is sent, the
power supply output voltage is set to 0 during the time that the relays are changing polarity.
 
 If the power supply is not configured for the relay option, sending either relay command generates an error.
 
 
Command Syntax 
OUTPut:RELay:POLarity   <CRD>
 
Parameters 
NORMal | REVerse
 
*RST Value
NORM
 
Examples
OUTP:REL:POL  NORM
 
Query Syntax
OUTPput:RELay:POLarity?
 
Returned Parameters 
NORM | REV
 
Related Commands
OUTP[:STAT]    *RCL    *SAV