46     Language Dictionary
TRIG:SOUR
 
 This command selects the trigger source.  Since the power supply has no other trigger source than the GPIB bus, this
command need not be used.  It is included in the command set to provide programming compatibility with other instruments
(such as the Agilent Electronic Load family) that may have more than one trigger source.
 
 
Command Syntax
TRIGer:SOURce <CRD>
 
Parameters
BUS
 
*RST Value
BUS
 
Examples
TRIG:SOUR  BUS      TRIGGER:SOURCE BUS
 
Query Syntax
TRIGger:SOURce?
 
Returned Parameters
BUS
 
Related Commands
*RST *TRG TRIG[:IMM]
 
Voltage Subsystem
 
 This subsystem programs the output voltage of the power supply.
VOLT
VOLT:TRIG
 
 These commands set the immediate voltage level or the pending triggered voltage level of the power supply.  The immediate
level is the voltage programmed for the output terminals.  The pending triggered level is a stored voltage value that is
transferred to the output terminals when a trigger occurs.  A pending triggered level is unaffected by subsequent VOLT
commands and remains in effect until the trigger subsystem receives a trigger or an ABORt command is given.  If there is
no pending VOLT:TRIG level, then the query form returns the VOLT level.  In order for VOLT:TRIG to be executed,
the trigger subsystem must be initiated (see INITiate).
 
 
Command Syntax 
[SOURce]:VOLTage[:LEVel][:IMMediate][AMPLitude] <NRf+> 
[SOURce][:VOLTage[:LEVel]:TRIGgered[:AMPLitude] <NRf+>
 
Parameters
Table 3-1
 
Default Suffix
V
 
*RST Value
Table 3-1
 
Examples
VOLT 200 MV        VOLTAGE:LEVEL 200 MV
 
VOLTAGE:LEVEL:IMMEDIATE:AMPLITUDE 2.5
 
VOLT:TRIG 20        VOLTAGE:LEVEL:TRIGGERED 20
 
Query Syntax
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]? 
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]? MAX
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?  MIN 
[SOURce]:VOLTage[LEVel]:TRIGgered[:AMPLitude]? 
[SOURce]:VOLTage[LEVel]:TRIGgered[:AMPLitude]?  MAX 
[SOURce]:VOLTage[:LEVel]:TRIGgered[:AMPLitude]?  MIN
 
Returned Parameters
<NR3> VOLT? and VOLT:TRIG? return presently programmed 
immediate and triggered levels.  If not triggered level is programmed, 
both returned values are the same.
 
VOLT?  MAX and VOLT?  MIN return the maximum and 
minimum programmable immediate voltage levels.
 
VOLT:TRIG?  MAX and VOLT:TRIG? MIN return the 
maximum and minimum programmable triggered voltage levels.
 
Related Commands
For VOLT     *SAV     *RCL      *RST
 
For VOLT:TRIG       ABOR     VOLT     *RST