Adam Equipment qbw Manuale Utente

Pagina di 16
© Adam Equipment Company 2004 
 
12 
 
  Factory defaults will be loaded for the stored setup, zero and calibration 
values, then the raw ADC counts will be displayed as described above. 
 
  This  mode  can  be  exited  at  any  time  by  pressing  the  [Cal]  key,  at 
which point, since the zero and calibration values have been defaulted, 
the  scale  will  automatically  enter  a  calibration  routine  as  described 
above.  A  calibration  procedure  must  then  be  carried  out  to  ensure 
reading accuracy. It is therefore vital that a standard calibration mass is 
available for this purpose. 
 
  A  successful  calibration  at  this  point  will  return  the  scale  to  normal 
weighing  mode,  but  any  errors  will  be  unrecoverable,  and  the  scale 
must be switched  off and back on again to allow the calibration to be 
completed.  Normal  weighing  will  not  be  possible  until  a  successful 
calibration has been performed after loading the default values. 
 
9.0  ERROR MESSAGES 
 
During the operation the software will attempt to ensure that the results obtained are 
sensible.  Potential  error  messages  and  their  possible  causes  are  detailed  in  the 
following table: 
At power-up 
A/D counts more than 10% of scale range below the default no-load value. 
A/D counts more than 10% of scale range above the default no-load value. 
Taring 
 
Taring could not be performed because the readings were not stable. 
No-load Calibration 
A/D counts more than 10% of scale range below the current no-load value. 
A/D counts more than 10% of scale range above the current no-load value. 
 
No-load value could not be determined because the readings were not 
stable. 
Full-load Calibration 
A/D counts more than 10% of scale range below the stored full-load value. 
A/D counts more than 10% of scale range above the stored full-load value. 
 
Full-load value could not be determined because the readings were not 
stable. 
Weighing 
!
Weight is more than 0.03% above the maximum range. 
!
Result exceeds the range of display.