2.  Locate the two mounting holes for the handle 
assembly on the Mini-Guardian HEPA System cabinet. 
They are located in close proximity to the top of the 
cabinet near the corners. A lock nut should be visible 
through each of the mounting holes and is held in 
place by a retainer plate secured to the cabinet shell. 
Verify lock nut is visible in each mounting hole.
3.  Place a socket head cap screw (M0 x 35mm)  
into each of the holes in the handle end brackets.  
Position the handle assembly up against the  Phoenix 
Mini-Guardian HEPA System cabinet so as to align the 
cap screws with the mounting holes in the cabinet.
 
It may be necessary to tilt the handle assembly  
up at a slight angle in order to align the screws  
with the lock nuts. Use the 8mm hex key to press  
the cap screws into the mounting holes and engage 
the threads of the lock nuts.
4.  Tighten each of cap screws using the 8mm hex key  
until the handle assembly is secured against the  
cabinet surface and all space between parts is  
eliminated.
2.2 Transporting the Phoenix
The Phoenix Mini-Guardian HEPA System should be upright 
when transported by vehicle. It may be tipped on to its 
back for loading and moving by hand or for use on location.
2.3 Location
Note the following precautions when locating the Phoenix 
Mini-Guardian HEPA System:
·  It is designed to be used INDOORS ONLY.
·  If used in a wet area, plug it into a GROUND FAULT 
   INTERRUPTER.
·  DO NOT use the Phoenix Mini-Guardian HEPA System as 
a bench or table.
·  The air inlet on top & the side outlets should be at  
 
least  foot from walls and other obstructions to  
 
airflow.
2.4 Electrical Requirements
The Phoenix Mini-Guardian HEPA System can be plugged 
into a grounded 5 Amp circuit. It draws about 2 Amps 
with clean filters and no ducting. The amp draw will 
decrease slightly with added ducting and as the filters load 
with particulate.
If an extension cord is required, it must have a minimum of 
8 gauge conductors if 25 feet long or less and 6 gauge 
conductors if greater than 25 feet long.
Figure 4: Tightening screws
Figure 3: Handle tilting for assembly
Figure 2: Mounting holes