The Calibration is complete, the scale 
will run the self-test during which the 
weight should be removed.  
 
If calibration is not possible an error 
message will be shown 
 
“- - - -”    “  - - - -”    “ - - - - ” 
“SPAn”    “ PASS”    “        ” 
 
  
“SPAn”    “ FAiL”    “ LoW   ”  
or 
 
“HiGH 
 
If The Calibration is being done while in the TECH parameters (section 12)  the linearity is 
set at the same time as calibration by performing the calibration in 2 steps,   
 
Display will ask to unload any weight on 
the platform. 
Press the [Print] key to continue. 
“nO L o Ad” 
 
 
 
   
Load the first calibration weight onto the 
scale  
 
If the mass is different than the value 
shown press the [CE] key then enter 
the calibration weight to be used . 
 
Then press [Print] to continue or press 
[Zero] to complete with only one mass 
being used. 
 
“LoAd  1 ”    “ 0010”    “ KiLoS”       
 
 
 
Load the second  calibration weight 
onto the scale  
 
If the mass is different than the value 
shown press the [CE] key then enter 
the calibration weight to be used . 
 
Then press [Print]. 
 
“LoAd  2”    “ 0030”    “ KiLoS”       
 
 
 
The Calibration is complete, the scale 
will run the self-test during which the 
weight should be removed.  
 
“- - - -”    “  - - - -”    “ - - - - ” 
“SPAn”    “ PASS”    “        ” 
 
If calibration is not possible an 
error message will be shown 
“SPAn”    “ FAiL”    “ LoW   ”  
or 
 
“HiGH