Adam Equipment mdw-300l Manuale Utente

Pagina di 20
© Adam Equipment Company 2012
 
           
Page 
9
 
 
 
6.0  CALIBRATION 
 
Before calibrating the scale, you should ensure that you have a suitable known 
weight for calibration.  
 
1.  When  in  normal  weighing  mode  with  the  scale  at  zero  press  and  hold  down 
[TARE] and [ON / OFF] keys to enter the calibration mode.  
 
2.  If the calibration switch is in the off position on the main PCB inside the scale, the 
indicator will show “
CAL.OFF
” and then exit this mode. If the indicator shows “C
AL-
?
”, the scale is ready for calibration. 
 
3.  When  the  indicator  shows  “C
AL-?
”,  press  the  [TARE]  key  to  confirm  and  go  to 
next step, or press the [ON / OFF] key to exit the calibration mode.  
 
4.  When  ‘
0.0
0.0
0.0
0.0
’  is  displayed  the  scale  will  begin  to  calibrate  the  scale’s  zero-point. 
Ensure that there is no load or weight on the scale’s platform. Press the [TARE] 
key to confirm, or press the [ON / OFF] key to exit this mode.  
 
5.  A  few  seconds  after  the  [TARE]  key  has  been  pressed  in  step  4  the  scale  will 
show “300.0” and the kg LED, or “600.0” and the lb LED depending on which unit 
you  chosen,  this  is  the  default  calibration  weight  from  the  factory.  Press  the 
[UNIT] key to select the calibration weight unit that you want to calibrate in. Press 
the  [ON  /  OFF]  key  to  exit  the  calibration  mode  at  this  point  or  press  the 
[Print/Hold]  key  to  choose  a  different  calibration  weight  value  (50kg,  100kg, 
150kg,  200kg,  250kg,  300kg  or  100lb,  200lb,  300lb,  400lb,  500lb,  600lb);  Then 
put  on  the  weight  that  you  selected  and  press  the  [TARE]  key  to  confirm  the 
chosen  standard  weight  that  was  selected  earlier.  The  displayed  data  will  flash 
on  the  display  and  if  the  scale  accepts  the  calibration  data  it  will  calculate  and 
store  the  information  into  the  EEPROM.  If  an  error  has  occurred,  the  scale  will 
display “CAL. Er” and return back to step 4 for re-calibration. If the loaded weight 
is  not  within  the  range  of  95%  to  105%  of  the  weight  value  you  selected,  the 
scale  will  not  calibrate  but  display  “CAL.  Er”  and  return  back  to  step  4  for  re-
calibration. ) 
 
6.  Check  the  calibration  by  putting  the  weight  that  you  calibrated  at  back  on  the 
scale, if it is off repeat the calibration process again.