Adam Equipment pgl series Manuale Utente

Pagina di 40
 
15 | 
P a g e  
 
© Adam Equipment Company 2011
 
12)
 
Place unknown sample to display the percentage weight 
13)
 
To  perform  percent  weighing  with  another  sample  press  [Mode]  and 
continue from step 2 above. 
14)
 
To return to normal weighing, press [Mode] then press [Esc]. 
Note: Percentage will be displayed to the maximum number of decimal places based 
on  the  resolution  of  the  balance.  To  increase  or  decrease  by  one  decimal  place, 
press the [Up] or [Down] key respectively. 
7.3 
CHECK WEIGHING 
During weighing of a sample the balance can be set to show if the weight is above or 
below  an  upper  and  a  lower  limit.    The  display  will  use  the  arrows  under  the  fill 
guide to show the check weighing is operating.   The arrows and bars between the 
arrows will indicate when the weight is below the lower limit, between the limits or 
above the upper limit. 
 
 
The buzzer can be set to be active when the weight is outside the limits (below the 
lower or above the upper) or within the limits (above the lower and below the upper 
limit), or turned off.  If desired, only one limit needs to be set. If only one limit is set 
the other limit is considered to be zero (lower) or the maximum (upper). 
The Check weighing is not active from zero to 20 scale divisions however the arrows 
will still be turned on below the fill guide and the weight will still be displayed. 
 
Steps: 
1)
 
Press [Mode] and then the [Up] or [Down] key to select Check Weighing.  When 
“CHECK” is displayed press the [Enter] key to enter Check Weighing mode 
2)
 
Press [Up] or [Down] to set the “LOW LIMIT” to “On” or “OFF” 
3)
 
Press  [Enter]  to  proceed.    If  the  “LOW  LIMIT”  was  set  to “On” the  display  will 
show the current low limit and allow you to change the limit using the numeric 
entry  method  (see  section  5.1).  The  display  will  show  the  weigh  in  the  current