10.0  BATTERY
 
OPERATION
 
The scale may operate from a built in lead acid battery.   During operation with the 
battery the backlight should be disabled for best battery life. Typical battery life 20 
hours. 
If  possible  disable  the  backlight  automatically  if  the  external  power  supply  is  not 
turned on. 
11.0  ERROR
 
CHECKING 
During weighing the balance is constantly checking to see if the balance is operating 
within the limited parameters.  The errors likely to occur are: 
 
A/D counts below lowest allowed value 
 
A/D counts above highest allowed value 
 
A/D not operating 
 
Maximum capacity exceeded 
Other errors may be detected during special functions or operations.  These will be 
described in the section that applies. 
Error messages and the reasons are: 
Concerning calibration 
ERROR StAb 
Calibration could not be completed  
because the results were not stable  
ERROR LOW or  ERROR HIGH 
Calibration constant not within 20% of old calibration 
constant 
Concerning weighing 
ERROR LOW 
Weight display is below zero by >4%max 
ERROR HIGH 
Weight is above maximum plus 90d 
ERROR ZERO 
Weight  on  pan  exceeds  zero  limit  when  power  is 
applied.