The “COMMAND” pot adjusts the Analog command signal sent to the flow controller.  The setting for 
each controller connected to the power supply can be observed. (Depending on how the power supply 
was set up, the display could indicate in flow units or percent of full scale). 
 
Hastings Power Supply Feature Guide 
Feature Model 
40  Power Pod 100  Power Pod 400 
Digital Readout 
9
 
9
 
9
 
±15 Volts 
9
 
9
 
9
 
Analog Outputs 
9
 
9
 
9
 
Controller  
9
 
9
 
Analog Control 
9
 
 
 
Front Panel Override 
9
 
 
9
 
Totalizer  
 
9
 
4 -20 mA 
9
 
 
9
 
Ratio Control 
 
 
9
 
Alarms 
9
 
9
 
9
 
Multi-Channel Display 
 
 
9
 
Conversion Factors 
9
 
9
 
9
 
Communications  
RS232 
RS232/RS485 
 
2.7.2.  Operation with a Power Supply other than a Hastings 
The flow controller must be connected to the power source as specified in section 2.6.  In general, a 0 - 
5 VDC command signal proportional to the intended flow (0 Volts = zero flow; 5 Volts = 100% of rated 
flow) must be applied to pin 14 of the DA-15 (15 Volt), Pin 3 of the DE-9 (24 Volt) connector.  A 0 - 5 
VDC signal proportional to the flow rate through the instrument will be present on pin 6 of the DA-15 
(15 Volt), Pin 2 of the DE-9 (24 Volt) connector.  The control mode is selected via pin 8 of the DA-15 
(15 Volt), Pin 6 of the DE-9 (24 Volt) connector.  Apply > 10 Volts for full open, < 0 Volts for closed 
and allow the override to float for flow proportional to the command voltage.  Refer to your power supply 
manual for the specifics of implementing these parameters.