SECTION 6   GPIB
6-54
100  ’ *************************************************
110  ’  *
*
120  ’ *         INPUT vs OUTPUT characteristic of amplifier
*
130  ’ *                            measurement program
*
140  ’  *
*
150  ’ *************************************************
160  ’
170  ’ ************** initialize GPIB system ****************
180  ’
190  ISET IFC’
........................................................................................................ Sends interface clear
200  ISET REN’
................................................................................................ Makes remote enable true
210  CMD DELIM=2’
.................................................................................................... Sets the delimiter to LF
220  CMD TIMEOUT=5’
............................................................................... Sets the time-out parameter to 5 sec.
230  ’
240  ’ *********** input measurement parameter  *************
250  ’
260  INPUT “MEASUREMENT FREQUENCY (MHz)” ;MF
270  ’
280  ’ ************** setting the MG3641A/MG3642A *****************
290  ’
300  PRINT @3; “*RST” ’ ............................................................................... Initializes the MG3641A/MG3642A
310  PRINT @3; “:FREQ”+STR$(MF)+“MHz” ’ ............................................................................. Sets a frequency
320  PRINT @3; “:AMPL:STAT OFF” ’ ..................................................................................... Turns off the output
330  ’
340  ’ ************** setting the MS2651A *****************
350  ’
360  PRINT @1; “INI” ’
.................................................................................................. Initializes the MS2651A
370  PRINT @1; “SP 100KHZ” ’ ......................................................................... Sets the frequency span to 100 kHz
380  PRINT @1; “CF”+STR$(MF)+“MHZ” ’ ........................................................................ Sets a center frequency
390  PRINT @1; “RL 20DBM” ’ .......................................................................... Sets the reference level to 20 dBm
400  ’
410  ’ ****************** measurement  *******************
420  ’
430  PRINT “INPUT OUTPUT GAIN”
440  PRINT @3; “ :AMPL:STAT ON” ’ ...................................................................................... Turns on the output
450  FOR LEVEL1  = –30 TO 0 STEP 2
460  PRINT @3; “ :AMPL:LEV”+STR$(LEVEL1)+“dBm” ’ .................................................... Sets the output level
470  FOR TIMER = 0 TO 1000:NEXT TIMER
480  PRINT @1; “TS;MKPK” ’ ........................................................................ Sweeps once and searches for a peak
490  PRINT @1; “MKL?” ’
.................................................................. Issues a marker level reading data request
500  INPUT @1;LEVEL2
510  PRINT USING “###.#dBm   ###.#dBm   ###.#dB”;LEVEL1,LEVEL2,LEVEL2-LEVEL1
520  NEXT LEVEL1
530  ’
540  END