9.0  ADJUSTMENT AND REPAIRS 
 
 
9.1  ADJUSTMENT 
 
The users can not make any adjustments other than Calibration. 
 
 
9.2  REPAIRS 
 
9.2.1  ELECTRONICS REPAIR 
 
The most common electrical problems are-  
 
1> Broken cables between  the base and indicator, 
2> Broken cables or poor connections on cables within the indicator. 
3> Lead Acid battery has reached the end of its life. 
 
Check the cables for loose connections, poor contact or abrasions on the sides and 
replace as necessary. 
 
The battery for the scales is a Lead Acid type with 6V, 5AH capacity.   If the battery 
must be replaced, use another Lead Acid battery of similar capacity only. 
 
If  the  battery  shows  less  than  6.5Vdc  after  charging,  it  is  probably  not  holding  the 
charge and should be replaced. 
 
The electronics used in the MDW- 250L scales consist of a main PCB assembly with 
a second small PCB for the keys.   The main PCB assembly is not user serviceable.  
If a problem is found on the PCB, it should be replaced. 
 
Contact your Dealer or Adam Equipment for assistance. 
 
 
9.2.2  MECHANICAL REPAIR 
  
 
The mechanics of the scale consists of a lever system to transfer the weight to the 
single load cell.   The lever system has series of bearings and knife edges to transfer 
the weight from the platform to the load cell.