Adam Equipment mdw-250 l Manuale Utente

Pagina di 16
© Adam Equipment Company 2005 
 
•  Overload  display:  When
  “
FULL
” 
appears  on  the  display,  it  shows  that 
the  load  on  the  platform  is  over  the  maximum  capacity.  Under  these 
circumstances, it is necessary to reduce the load otherwise the sensor 
or the platform will be damaged. 
  
•  Hold Function: To lock the weighing result, press the [Hold] key. The 
arrow  indicator  will  point  towards 
HOLD  and  it  will  the  show  a  steady 
reading even if the person on the platform is moving or goes away from 
the scale and will not affect the next weighing. To release the function, 
press the 
[Hold] key again. 
 
•  Do not press the [On/Off] key if you do not want to weigh as pressing 
the 
[On/Off] key will switch the scale to ON
 
6.2  MEASURING HEIGHT  
 
•  While  measuring  the  height,  it  is  necessary  to  pull  up  the  measuring 
board at a right angle with the inside tube.  
 
•  When  the  tube  is  pulled  out  straight,  it  is  sufficient  to  measure  the 
height form 70-126cm. The number can be obtained upon the conjoint 
place where the upper part of middle tube screw meets with the inside 
tube scale.  
 
•  Further,  if  the  middle  tube  is  pulled  out  straight,  it  is  possible  to 
measure  the  height  from  126-190cm.  The  reading  can  be  obtained  at 
the  conjoint  place  where  the  upper  part  of  outside  tube  screw  meets 
with the middle tube scale. 
 
 
7.0  USER CALIBRATION 
 
 
1.  Press 
[On/Off] key when the display shows
 ”
SCALE
”.  
 
2.  Press 
[On/Off] key again. 
 
 
If  the  display  is  not  showing
  “
Id01
” 
open  the  upper  housing  and  switch  the
 
span jumper on PCB to 
ON.  
 
Start again by pressing 
[On/off]
 
 
The display shows
 “
Id01
”, “
Id-02
”, “
Id05
”, “
Id10
”,”
Id-20
”, “
Id-50
”. 
 
 
3.  Select
 “
Id01
” 
for 100g increments and press the 
[On/Off] key once.