User Calibration Guide, 457-0931 rev 1.0 
  
Equipment 
Refer to the appropriate block below, depending on whether you will be calibrating 
through a 37-pin, 68-pin, or 100-pin connector.    
P1                       
 
                                                             37-PIN 
If calibration signals will be passing through a 37-pin P1 connector you will need:
 
Required: 
6.5-digit, digital multi-meter
1
adjustable voltage calibrator
2
ambient temperature meter 
In addition, the following are recommended
DBK11A, passive screw-terminal board 
CA-37, 37 conductor ribbon cable 
 
P4                                                                                   100-PIN 
If calibration signals will be passing through a 100-pin P4 connector you will need:
 
Required: 
6.5-digit, digital multi-meter
1
adjustable voltage calibrator
2
ambient temperature meter 
In addition, the following are recommended: 
DBK11A, passive screw-terminal board* 
DBK200, P4-to-DB37 adapter card* 
CA-195 100-conductor ribbon cable 
*Another DBK200 Series device that includes a P1 and P4 connector may be substituted. 
  See page iv. 
 
P5   SCSI                                                                          68-PIN 
If calibration signals will be passing through a 68-pin SCSI (P5) connector you will need:
 
Required: 
6.5-digit, digital multi-meter
1
adjustable voltage calibrator
2
ambient temperature meter 
In addition, the following are recommended
TB-100, passive screw-terminal board 
CA-G56 (3 ft. shielded cable) ** 
** One of the following cables may be substituted: CA-G55, a 3 ft. unshielded cable; or 
     CA-G56-6, a 6 ft. shielded cable. 
 
Certain steps in the calibration process require that you apply a signal to the device via a specific pin, or 
take a measurement reading from a specific pin.  To make the task easier we recommend that you use a 
passive screw-terminal board, such as the DBK11A, TB-100, or possibly a DBK200 Series device.   
See the Setup section beginning on page iii.  
 
1
Agilent 34401A or better DMM.
 
2
Traceability is through the DMM.  The calibrator is only used as a quiet and stable voltage source.