Hanna Instruments lp_2000 Manuale

Pagina di 16
3
Remove the instrument from the packing material and examine it to
make sure that no damage has occurred during shipping. If there is
any damage, notify your dealer immediately.
LP 2000 is supplied complete with:
• Measurement cuvette and cap
• HI 93703-0 AMCO-EPA-1 calibration solution @0 FTU (30 mL)
• HI 93703-10 AMCO-EPA-1 calibration solution @10 FTU (30 mL)
• 12VDC Transformer (HI 710005 or HI 710006)
Note: Save all packing material until you are sure that the instrument
functions correctly. Any damaged or defective item must be
returned in its original packaging with the supplied accessories.
PRELIMINARY EXAMINATION
LP 2000 is a bench microprocessor-based meter used to determine
the turbidity of water and wastewater. The meter covers the 0 to
1000 FTU* range with 0.01 resolution from 0.00 to 50.00 FTU* and
in single digits from 50 to 1000 FTU*. It also auto-ranges between
the two scales.
Note: LP 2000 has been designed according to the ISO 7027
International Standard. The turbidity measurement unit
is FTU (Formazine Turbidity Unit), which is identical to
NTU (Nephelometric Turbidity Unit).
LP 2000 carries the 
 mark, in compliance with the European
directives, ensuring low emission values and accurate measurements
in presence of electromagnetic fields.
The meter is housed in a rugged and lightweight case, with an easy-
to-read LCD.
The meter is simple to use. All operations can be carried out with only
four keys and troubleshooting functions can be performed with error
codes on the display. An exclusive positive-locking system guarantees
that the cuvette is firmly placed in the cell. The keypad is water-
resistant and can be wiped clean.
Calibration at 10 FTU* can be easily performed using the supplied
standard. In addition, LP 2000 is the first bench turbidity meter
that allows the storage and retrieval of the last calibration date.
10 FTU* was chosen as the calibration point because its value best
fits the water turbidity measurements in a wide range of applications,
from drinking water to wastewater treatment.
GENERAL DESCRIPTION
* 1 FTU = 1 NTU