Chapter 1
Installing and Configuring the SCXI-119X
1-12
ni.com
e.
Change the Connected to selection from None to Connected to:
Device x
.
If you get the warning message
Failed to find
followed by the
module codes and the message
Unable to communicate with 
chassis
, take the following troubleshooting actions:
a.
Make sure that the SCXI chassis is powered on.
b.
Make sure the cable between the SCXI chassis and DAQ device or
computer-based instrument is properly connected.
c.
Inspect the cable and the connectors for any bent pins.
d.
Make sure you are using the correct National Instruments cable
assembly.
e.
Test your DAQ device or computer-based instrument to verify it is
working properly. Refer to your DAQ device or computer-based
instrument user manual for more information.
If you get the warning message
Failed to find
, followed by
module codes and then the message
Instead found: module with 
ID 0Xxx
:
a.
Return to the
section and make sure the correct
module is in the specified slot.
b.
Delete the incorrect module as described in the
section.
c.
Add the correct module as described in the
section.
If you get the warning message
Failed to find
, followed by a
module code and then the message
Slot x is empty
:
a.
Check to see if the configured module is installed in the specified
slot. If not, install the module by referring to the
section.
b.
If the module is installed in the correct slot, turn off the chassis,
and remove the module as specified in the
section. Verify that no
connector pins are bent on the rear signal connector. Reinstall the
module, ensuring the module is fully inserted and properly aligned
in the slot.
After checking the preceding items, return to the
section and retest your SCXI chassis.