NOTE FOR USERS OF NEW
molbox1™
WITH EXISTING molblocs™
If you are going to be using this new molbox1 with previously delivered molblocs, there is the possibility that
the molbloc temperature measuring circuit is not compatible with the new molbox1.
To check compatibility, turn on the molbox1 with the molbloc connected. If the molbox1 autotests and goes
to the main run screen normally, the molbloc is compatible with the molbox1. If you encounter the message
"PRT ERROR", the molbloc is not compatible with the molbox1. If used, it may give out of tolerance
flow indications.
If you observe a "PRT ERROR" message when initializing a molbloc, contact DHI Technical Service for
information on upgrading the molbloc for compatibility.