8  • Error Messages Troubleshooting and Diagnostics 
RX Only –NGP Z370500 Rev. B 1116 
8-38 
 
 
 
 
513 IV Pole communications error. 
Probable cause: Communication error
Corrective action
1.   Prime the system. 
 
2.   If the error still does not clear or if the error continues to occur, document the 
error message and contact AMO for technical service. 
 
514 IV Pole error. 
 
Probable cause: Communication error
.
 
 
Corrective action
 
1.   Prime the system. 
 
2.   If the error still does not clear or if the error continues to occur, document the 
error message and contact AMO for technical service. 
 
515 IV Pole calibration error. 
Probable cause: Communication error
Corrective action
1.   Run Self Test
 
2.   If the error still does not clear or if the error continues to occur, document the 
error message and contact AMO for technical service. 
 
516 IV Pole communication error. 
Probable  cause:  Communication error. 
Corrective action
1.   Prime the system. 
 
2.   If the error still does not clear or if the error continues to occur, document the 
error message and contact AMO for technical service.