Introduction
x
Contraindications
Insulin pump therapy is NOT recommended for people who are 
either:
Unable to perform at least four (4) blood glucose tests per day
Unable to maintain contact with their healthcare provider
Unable to use the System according to instructions
Do not use the integrated FreeStyle® Blood Glucose Monitoring 
System for:
Testing on newborns
Testing arterial blood
Diagnosing of or screening for diabetes mellitus
The OmniPod System is designed to use rapid-act-
ing U-100 insulin.  The following U-100 rapid-acting 
insulin analogs have been tested and found to be 
safe for use in the Pod:  Novolog®/NovoRapid®, 
Humalog®, or Apidra®. Novolog® is compatible with 
the OmniPod System for use up to 72 hours (3 days). 
Before using a different insulin with the OmniPod 
System, check the insulin drug label to make sure it 
can be used with a pump. Refer to the insulin label-
ing and follow your healthcare provider’s directions 
for how often to replace the Pod.
Not recommended for individuals with hearing 
loss. Always verify ability to hear Pod/PDM alerts 
and alarms.
The Pod and PDM may be affected by strong radi-
ation or magnetic fields. Before having an X-ray, MRI, 
or CT scan (or any similar test or procedure), remove 
your Pod and put it and the PDM outside the treat-
ment area. Check with your healthcare provider for 
Pod removal guidelines.
The glucose meter and lancing device are for single 
patient use. Do not share the device with anyone 
including other family members. Do not use on mul-
tiple patients.
Parts of the device are considered biohazardous and 
can potentially transmit infectious diseases, even 
after you have performed cleaning and disinfection.