Insulet Corporation 017 Manuale Utente

Pagina di 190
92
7
■ 
Entering Blood Glucose Readings Manually
You can enter blood glucose readings manually into the PDM. 
This is particularly helpful if you use a separate blood glucose 
meter. 
1. On the Home screen, use the Up/Down Controller buttons 
to choose More actions,then press Select.
2. Choose Add BG reading, then press Select.
3. Use the Up/Down Controller buttons to enter the BG read-
ing from your separate meter.
4. Press Tag to assign a tag to the reading. Press OK when fin-
ished. (For an explanation of entering tags, see page 90.)
5. Press Save to save the reading.
Or press Cancel to return to the More actions menu without 
saving the reading.
The Bolus Entry screen does not open if:
A regular bolus is currently in process
The meter temperature is out of range
You have tagged this BG reading as a Control
The BG reading is less than your “Min BG for calcs”
Insulin is suspended
Figure 7-33
To enter a “LOW” reading, press the Up/Down 
Controller
 buttons down until you reach the mini-
mum value (20); then press down one more time. 
To enter a “HIGH” reading, press up until you reach 
the maximum value (500); then press up one more 
time.