The PDM automatically alerts you when its battery power gets 
low (see Figure A-1). When you see the “Low PDM battery” alert, 
press OK to turn off the alert. Replace the batteries as soon as 
you can.
To replace batteries
1. Open the battery compartment on the back of the PDM by 
pushing the compartment latch inward and then upward 
(Figure A-2). No special tools are necessary.
2. Insert 2 new AAA alkaline batteries into the compartment.
3. Replace the battery door.
4. Turn the PDM face up. The PDM powers on automatically.
The PDM is now ready to use. Dispose of the old batteries in 
accordance with local waste disposal regulations.
If the PDM becomes submerged in water
The PDM is not waterproof. Do not place it in water or leave it 
near water where it can accidentally fall in. If it becomes sub-
merged in water:
1. Dry the outside of the PDM with a clean, lint-free cloth.
When your PDM alerts you that battery power is getting 
low, the PDM will maximize remaining battery life by:
Turning off your vibration alert (if set)
Disabling bright mode
Disabling the test strip port light
Once the batteries are replaced, these functions resume.
Figure A-1
Figure A-2
If the batteries run out, data in the memory is at risk. 
DO NOT remove the old batteries until you have 
new ones at hand. The PDM protects data in the 
memory for up to 2 hours after the batteries run out 
or are removed.
The diagram inside the compartment shows you 
which direction to insert the batteries.