Insulet Corporation 017 Manuale Utente

Pagina di 190
33
4
4. If you have set the blood glucose reminder option to On and 
you want to add a reminder:
a. Press Yes.
b. Enter the time for the reminder, then press OK.
5. Press Confirm to begin the bolus delivery.
Deliver a bolus manually (suggested bolus calc is Off)
1. On the Home screen, use the Up/Down Controller buttons 
to choose Bolus, then press Select.
2. Enter the number of units you want to deliver, then press 
Enter.
3. If the blood glucose reminders feature is turned On, you can 
now set a blood glucose reminder. 
4. Press Confirm to begin the bolus delivery.
If the reverse correction feature is turned On and if 
your blood glucose level is below your target, the 
System subtracts a correction amount from the meal 
portion of the bolus.
If the reverse correction setting is turned Off, the 
System does not subtract anything for a blood glu-
cose level below your target.
Figure 4-6
As a safety feature, the OmniPod System only allows 
you to give a bolus at or below the maximum bolus 
dose you have set. See Chapter 6, Using the Personal 
Diabetes Manager
, for information on resetting your 
maximum bolus dose. If suggested bolus calc is On
you can enter a temporary manual override to make 
an adjustment above your max bolus limit (Figure 4-1 
and Figure 4-2). Consult your healthcare provider 
before changing this setting.