Insulet Corporation 019 Manuale Utente

Pagina di 190
158
General Notes
Medical electrical equipment requires special precautions 
regarding electromagnetic compatibility (EMC) and needs to be 
installed and put into service according to the EMC information 
provided in this document and the instructions for use.
Portable and mobile RF communications equipment can affect 
the function of medical electrical equipment.
Insulet Corporation declares that the mylife OmniPod Insulin 
Management System is in compliance with the essential require-
ments and other relevant provisions of Directive 1999/5/EC.
Cables and accessories not specified within the instructions for 
use are not authorized. Using other cables and/or accessories 
may adversely impact safety, performance, and electromagnetic 
compatibility (increased emission and decreased immunity).
Care should be taken if the System is used adjacent to other elec-
trical equipment; if adjacent use is inevitable, such as in work 
environments, the System should be observed to verify normal 
operation in this setting.
The System communicates by low level RF energy. As with all RF 
receivers, the potential for disturbance exists, even with equip-
ment that complies with FCC and CISPR emissions requirements.
The mylife OmniPod System communicates with the following 
characteristics:
Frequency: 433 MHz, FSK modulation, with an effective radi-
ated power of 13mW
The mylife OmniPod System greatly exceeds the immunity 
requirements of the general standard for electromagnetic com-
patibility, IEC 60601-1-2.
Electromagnetic Emissions
This equipment is intended for use in the electromagnetic environment specified below. The user of this equipment should assure 
that is used in such an environment.
Emissions
Compliance according to
Electromagnetic environment
RF Emissions (CISPR 11)
Group 2
The Pod and the PDM emit low level electromagnetic energy (RF) in 
order to communicate. Although unlikely, nearby electronic equip-
ment may be affected.
CISPR B 
Emissions Classification
Class B
The System is suitable for use in all establishments including domestic 
establishments.