Insulet Corporation 019 Manuale Utente

Pagina di 190
33
4
   
4. If you have set the blood glucose reminder option to On and 
you want to add a reminder:
a. Press Yes.
b. Enter the time for the reminder, then press OK.
5. Press Confirm to begin the bolus delivery.
Deliver a bolus manually (suggested bolus calc is Off)
1. On the Home screen, use the Up/Down Controller buttons 
to choose Bolus, then press Select.
2. Enter the number of units you want to deliver, then press 
Enter.
3. If the blood glucose reminders feature is turned On, you can 
now set a blood glucose reminder. 
4. Press Confirm to begin the bolus delivery.
   
If the reverse correction feature is turned On and if 
your blood glucose level is below your target, the 
System subtracts a correction amount from the meal 
portion of the bolus.
If the reverse correction setting is turned Off, the 
System does not subtract anything for a blood glu-
cose level below your target.
Figure 4-6
As a safety feature, the mylife OmniPod System only 
allows you to give a bolus at or below the maximum 
bolus dose you have set. See Chapter 6, Using the 
Personal Diabetes Manager, f
or information on reset-
ting your maximum bolus dose. If suggested bolus 
calc is On, you can enter a temporary manual over-
ride to make an adjustment above your max bolus 
limit (Figure 4-1 and Figure 4-2). Consult your health-
care provider before changing this setting.