Insulet Corporation 019 Manuale Utente

Pagina di 190
55
5
Or press Back to return to the previous screen.
 
Check Pod Status
To check Pod status, if the PDM is turned Off, press and hold the 
Power button to display the ID screen, then press Confirm to 
display the Status screen. If the PDM is On, press Back until you 
reach the Status screen. (If you press and hold the Power but-
ton, you will turn off the PDM instead.) Or, from the Home 
screen, press Status.
The PDM automatically checks the status of the Pod. The PDM 
then displays the current insulin reservoir volume, PDM battery 
level, last blood glucose value, last bolus dose, active basal pro-
gram, and any alarm conditions. See Chapter 6, Using the 
Personal Diabetes Manager, f
or details of what you see on the 
Status screen.
   
If the Pod goes into hazard alarm condition and the PDM is 
unable to communicate with it to turn off the alarm, you can turn 
it off manually on the Pod. See the end of Chapter 10, Alerts and 
Alarms, f
or instructions.
NEVER inject insulin (or anything else) into the 
fill port while the Pod is on your body. Doing so may 
result in unintended or interrupted insulin delivery.
Figure 5-19
Figure 5-23
Once a Pod is activated and communicates with the 
PDM, it can only receive commands from that PDM, 
not from any other.