Reset the Date or Time
Occasionally, you need to change date and time settings (for 
example, to adjust for daylight savings time or after resetting the 
PDM). As a safety feature, you can change date and time settings 
only when the Pod is deactivated or when insulin delivery is sus-
pended (see Chapter 5, Using the Pod).
1. Suspend insulin delivery.
 
2. On the Home screen, use the Up/Down Controller buttons 
to choose Settings; then press Select.
3. Choose System setup, then press Select.
4. Choose Date/time, then press Select.
5. Choose either Date or Time, then press Edit.
Reset the time
1. To enter the current time, press the Up/Down Controller 
buttons. Press and hold the button to increase or decrease 
the time faster.
2. Press 12/24 hr to choose either a 12-hour or 24-hour clock, 
then press Enter (Figure 6-4).
3. Press Confirm to accept the new time (Figure 6-5).
If you are changing the Pod, you can reset the date or 
time without suspending by changing the date or 
time before activating the new Pod.