Insulet Corporation 019 Manuale Utente

Pagina di 190
67
6
buttons to choose a new start time, end time, and IC ratio. 
Press Next after each entry.
b. To add a segment, choose [add new], then press New
Use the Up/Down Controller buttons to choose a start 
time, end time, and IC ratio. Press Next after each entry.
c. When you have completed all time segments, press 
Done, then Save.
4. Correction factor—You can edit existing time segments and 
add segments, up to a total of 8.
a. To edit an existing correction factor time segment, choose 
the segment, then press Edit. Use the Up/Down Control-
ler
 buttons to choose a new start time, end time, and cor-
rection factor. Press Next after each entry.
b. To add a segment, choose [add new], then press New
Use the Up/Down Controller buttons to choose a start 
time, end time, and correction factor. Press Next after 
each entry.
c. When you have completed all time segments, press 
Done, then Save.
5. Reverse correction—Choose On or Off, then press Select.
6. Insulin action (duration) or Insulin on Board (IOB)—Use 
the Up/Down Controller buttons to choose a duration of 
insulin action, then press Enter.
Temp basal
Choose %U/hr, or Off to set the mode for temporary basal rates 
and presets, then press Select.
Extended
Choose %Units, or Off to set the mode for extended boluses, 
then press Select.
Bolus increment
Choose 0.050.100.50, or 1.00 units for bolus increments, then 
press Select.
Max bolus
Use the Up/Down Controller buttons to enter the maximum 
bolus dose you can take, then press Enter.
Max basal rate
Use the Up/Down Controller buttons to enter the maximum 
basal rate you can choose, then press Enter
Some settings have preset defaults, but all settings 
are changeable. See the Appendix for a list of Sys-
tem specifications, including preset default settings.