6. Configuring the Receiver: Other Than Keypad and Display 
P5E GNSS Reference Receiver User Guide                                    
 Page    38 
 
In this screen, you can set all data logging parameters, and 
determine whether the recording files will be affected by the FTP 
Push. The parameters are mainly as follows: 
➢  Auto Record: Select “Yes” or “No” to determine whether to log 
data when the Master Log Switch is ON.   
➢  Sample Interval: Select the observable rate from the dropdown 
list.   
➢  Store Location: Determine whether to store at internal storage 
or external storage.   
➢  Start Date: Set the start time of data log in UTC. Select “Yes” or 
“No” option below to determine whether to start logging from 
the set time. 
➢  Duration Time: Set the time interval of recording. 
➢  Storage: Set the storage limit of this thread.   
➢  Circulating Memory: Select “Yes” or “No” to determine 
whether to auto delete old files if the storage space is full.   
➢  Storage Format: The default format of recording files is HCN.   
➢  FTP Push: Decide whether to push the stored files to the FTP 
server of your choice. 
Click 
  to save the settings and back to the Log Settings 
screen. Also, users can click 
  to abandon the changed 
settings and back to Log Settings screen.   
NOTE – to delete the record data, you can remove all the eight record 
data by the total button clear all accounts. Also you can delete each 
record data by clear button under the selected thread.