Xplore Technologies Corp. EM75 Manuale Utente
18
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions (CISPR 11)
Group 1
The XSlate L10 Rugged Tablet PC uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
The XSlate L10 Rugged Tablet PC is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
its RF emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
The XSlate L10 Rugged Tablet PC is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
RF emissions (CISPR 11)
Class B
Harmonic emissions (IEC 61000-
3-2)
3-2)
Class A
Voltage fluctuations/ flicker
emissions (IEC 61000-3-3)
emissions (IEC 61000-3-3)
Complies
Notes:
None
Table 1-2: Guidance and manufacturer’s declaration – electromagnetic immunity – For all ME equipment and ME systems
The XSlate L10 and XPad L10 Rugged Tablet PC is intended for use in the electromagnetic environment specified below. The customer or the
user of the XSlate L10 and Xpad L10 Rugged Tablet PC should assure that it is used in such an environment.
user of the XSlate L10 and Xpad L10 Rugged Tablet PC should assure that it is used in such an environment.
Immunity test
Test level
Compliance level
Electromagnetic environment – guidance
Electrostatic discharge
(ESD) (IEC 61000-4-2)
(ESD) (IEC 61000-4-2)
±6 kV contact
±8 kV air
±8 kV air
±6 kV contact
±8 kV air
±8 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30 %.
floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrical fast
transient/burst (IEC 61000-
4-4)
transient/burst (IEC 61000-
4-4)
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV for input/output lines
Mains power quality should be that of a typical
commercial or hospital environment.
commercial or hospital environment.
Surge (IEC 61000-4-5)
±1 kV differential mode
±2 kV common mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
±2 kV common mode
Mains power quality should be that of a typical
commercial or hospital environment.
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines (IEC 61000-4-
11)
interruptions and voltage
variations on power supply
input lines (IEC 61000-4-
11)
<5 % UT
(>95 % dip in UT) for 0,5 cycle
(>95 % dip in UT) for 0,5 cycle
40 % UT
(60 % dip in UT) for 5 cycles
(60 % dip in UT) for 5 cycles
70 % UT
(30 % dip in UT) for 25 cycles
(30 % dip in UT) for 25 cycles
<5 % UT
(>95 % dip in UT) for 5 sec
(>95 % dip in UT) for 5 sec
<5 % UT
(>95 % dip in UT) for 0,5 cycle
(>95 % dip in UT) for 0,5 cycle
40 % UT
(60 % dip in UT) for 5 cycles
(60 % dip in UT) for 5 cycles
70 % UT
(30 % dip in UT) for 25 cycles
(30 % dip in UT) for 25 cycles
<5 % UT
(>95 % dip in UT) for 5 sec
(>95 % dip in UT) for 5 sec
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the XSlate L10 Rugged Tablet PC requires continued
operation during power mains interruptions, it is
recommended that the XSlate L10 Rugged Tablet PC
be powered from an uninterruptible power supply or
a battery.
NOTE: UT is the a.c. mains voltage prior to application
of the test level.
commercial or hospital environment. If the user of
the XSlate L10 Rugged Tablet PC requires continued
operation during power mains interruptions, it is
recommended that the XSlate L10 Rugged Tablet PC
be powered from an uninterruptible power supply or
a battery.
NOTE: UT is the a.c. mains voltage prior to application
of the test level.
Power frequency (
50/60 Hz) magnetic field
(IEC 61000-4-8)
50/60 Hz) magnetic field
(IEC 61000-4-8)
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
characteristic of a typical location in a typical
commercial or hospital environment.
Notes:
None