KROHNE FMCW10G52 Manuale Utente

Pagina di 67
03/2012 - 4001904901 - HB OPTIWAVE 5200 R01 en 
www.krohne.com 
47 
  OPTIWAVE 5200 C/F  
OPERATION 
6
 
 
 
 
 
 
Measurement type definitions 
 
Measurement type
 
Description
 
Available units
 
LEVEL
 
This is a display and an output function option. It is 
the height from the bottom of the tank to the surface 
of the liquid (Tank height - Distance).
 
m, cm, mm, in (inches), ft 
(feet)
 
DISTANCE
 
This is a display and an output function option. 
It is the distance from the face of the flange to 
the surface of the liquid.
 
m, cm, mm, in (inches), ft 
(feet)
 
CONVERSION
 
This is a display and an output function option. It 
gives the volume, mass or flow rate of the tank 
contents. This data is available if you prepare a 
volume, mass or flow rate table in configuration 
mode. For 
data on how to prepare the conversion table, 
refer to 
How to configure the device to 
measure volume or mass 
on page 68.
 
m3, L, gal (US gallons), ImpG (Imperial 
gallons), ft3, bbl (oil barrel), kg, t, Ston, 
Lton, m, cm, mm, in, ft, m3/h, ft3/h
 
ULLAGE CONV.
 
This is a display and an output function option. It 
gives the empty volume or remaining mass that 
can be put in   the tank. This data is available if you 
prepare a volume or mass table in configuration 
mode. For data on how to prepare the conversion 
table, refer to 
How to configure the device to 
measure volume or mass 
on page 68.
 
m3, L, gal (US gallons), ImpG (Imperial 
gallons), ft3, bbl (oil barrel), kg, t, Ston, 
Lton, m, cm, mm, in, ft, m3/h, ft3/h
 
OUTPUT I (mA)
 
The current output of the device.
 
mA
 
OUTPUT I (%)
 
The percentage of the current output. 0% = 4 
mA. 100% = 20 mA.
 
%
 
 
 
6.3  Configuration mode 
 
6.3.1  General notes 
 
Change the settings of your device in Configuration mode. Data about the menus is given on page 
58.  You can: 
 
•   Use the 1.0.0 INFORMATION menu to read  settings, device software versions and error records. For more data about the 
Information menu, refer to Table A. Info. 
•   Use the 2.0.0 SUPERVISOR menu  run diagnostic tests, set up a conversion table for volume, mass or flow rate measurement, 
change critical parameters for difficult process conditions, reset the device and  change basic parameters (tank height etc.), 
output settings, HART Address etc. For more data about the Supervisor menu, refer to Table 2. Supervisor. 
 
 
INFORMATION! 
It is not possible to enter the 3.0.0 SERVICE and 4.0.0 MASTER menus. These menus are for 
factory calibration and qualified service personnel only.