VM-550/VM-560
User Guide V1.0
Introduction
This guide describes the operation of the VM-550 and VM-560 pipe 
and cable locators. Both are general purpose locators used to detect 
buried pipe and cable services in a variety of situations.
The  operations  of  the  two  systems  are  very  similar.  The  main 
difference is in the operational frequencies:
•   VM-550 Frequencies: 50 or 60Hz, 8 kHz, 83 kHz
•   VM-560 Frequencies: 50 or 60Hz, 8 kHz, 480 kHz
The VM-550/VM-560 is supplied  with 1 watt transmitter with both 
direct connection and induction capabilities. An inductive clamp and 
LPC live separation filter are also available as optional accessories.
To make a direct connection to a pipe 
or a de-energized cable conductor or 
cable sheath.
To make a ground connection to 
complete a direct connection circuit.
To make an inductive coupling 
connection to a live cable or pipe.
Receiver Display Functions 
1
Battery Level Indicator 
2
Speaker Level Indicator 
3
Percentage Signal Levels
4
Selected Locate Frequency
5
Signal Level Indicator 
6
Sensitivity Setting Indicator 
Receiver Operational Controls 
1
2
3
4
5
6
7
8
9
ON/OFF (Long press until Power ON indicator is on, long
press again to turn off)
Power ON Indicator 
Low Output Power Selected
High Output Power Selected
Speaker
Frequency Select
Frequency Selected Indicators
Battery Cover Retaining Screws
Battery Housing Cover
Transmitter Operational Controls 
Accessories
1
2
3
4
Direct Connection Leads 
Ground Stake 
Signal Clamp 
Power Mode Operation
Switch on the unit by pressing the ON/OFF pushbutton. Allow a few 
seconds to switch on.
The  frequency  selected  as  shown  on  the  display.  If  this  is  not  the 
desired locate frequency (ie 50 or 60Hz) change as indicated below.
NOTE
The  battery  condition  as  indicated  on  battery  icon. 
Replace  batteries  if  necessary.  See  Changing  Batteries 
Section.
Changing the Locate Frequency
Press  and  hold  the  depth  measurement/frequency  selection 
pushbutton  until  the  frequency  menu  is  entered.  The  display  will 
show the present frequency selected in large numeric’s in the centre 
of the screen. Use the “+” or “–” pushbuttons to select the desired 
frequency.  Press  the  depth  measurement/frequency  selection 
pushbutton to re-enter the locate screen.
Locating a Cable in the Power (50/60Hz) Mode
1.    Hold  the  locator  vertically  in  the  area  that  is  required  to  be 
searched. Press the “+” or “-” pushbuttons to set the gain so that 
the bar graph reads approximately 50%.
2.    Hold the locator in front of you in the orientation shown below.
 
3.   Sweep the locator left to right along the suspected route of the 
cable. As the locator approaches the cable the meter reading will 
increase.  Pinpoint  the  position  by  detecting  the  largest  signal. 
Adjust  the  sensitivity  of  the  locator  by  pressing  the  “+”  or  “-” 
pushbuttons to keep the signal on scale.
4.   To confirm the direction of the cable, rotate the locator until the 
largest  signal  is  detected.  The  direction  of  the  cable  is  then 
directly ahead.
5.    Continue to locate the cable along the route.
6.   Depth measurements are not possible in the power (50/60Hz) 
mode, if pressed by accident it will show N/A.
WARNING
The power mode is used to detect signals radiating 
from cables or services that are carrying a 50 or 
60Hz load. It is possible for a cable to be live but 
not carry a load. In this case there may not be a 
signal to be detected.
Similarly, if a cable is exactly balanced the 
resulting signal radiating from the cable may be 
zero and therefore not detectable.
Do NOT use the VM-550/VM-560 to identify if cables 
are live. Always dig with caution.
Active Cable and Pipe Locating
Detecting a cable or pipe can be achieved by applying a locate tone 
to a cable or pipe from a transmitter. This is called active locating.
The locate tone can be applied by either.
•      Direct Connection
•      Induction Mode
•      Signal Clamp Mode
Direct Connection Mode
This method involves making a direct connection to the Cable or Pipe.
WARNING
DO  NOT  attempt  to  make  a  connection  to  a  live 
conductor. Only make a connection to de-energized 
or dead cables. It is possible to connect to the sheath 
of active cables but this should only be attempted by 
qualified and authorized personnel.
Method:
Plug the direct connection leads to the transmitter. Connect the red 
lead  to  the  cable  or  pipe  and  the  black  one  to  a  suitable  ground. 
Ideally this should be a ground stake placed at right angles to the 
probable route of the target line. If it is not possible to use a ground 
stake, connect the black lead to a grounded structure such as the rim 
of a manhole cover or other buried metallic structure. Try to avoid 
fencing as this will create interference from the return signal travelling 
along the fence.
If necessary clean the connection point with a wire brush to ensure a 
good electrical connection.
Select  the  desired  frequency  by  pressing  the  “f”  pushbutton.  As  a 
general  rule  the  higher  frequency  will  have  a  cleaner  more  stable 
reading  and  will  jump  insulation  joints  on  pipes  but  has  the 
disadvantage that it is more likely to jump to other utility nearby. The 
lower frequency is better for tracing a particular utility as it is more 
likely to keep to the utility line the transmitter is attached to. If in doubt, 
start with the lower frequency and switch to the higher one if it is not 
possible to detect a stable reading.
The same applies to setting the signal level. A short press on the
61
83k
1
2
3
4
5
6
To safely inject a trace tone to a live cable 
via a domestic mains socket, so that the 
cable can be traced from the premises to 
the connection in the street.
LPC Separation Filter 
8
9
7
1
2
3
4
6
5
Sensitivity Control 
(reduce sensitivity)
Increment sensitivity down, or auto scale 
down to 50% if off scale. When in 
frequency select menu use this button to 
scroll backwards through available 
frequencies.