Troubleshooting
Sensor Belt
209
Sensor Belt
Capsule
Cradle
PillCam Recorder DR3
Problem
 Cause
Action
Connector is damaged
Mishandling the sensor 
belt
End of Life
Stop using this sensor
Contact Customer Support
Insulation of the sensor wire 
is damaged
Problem
 Cause
Action
DOA (Dead On Arrival)
Capsule failure
1 Send capsule to Given 
Imaging Ltd.
2 Open another capsule.
3 If second capsule from 10-
pack is DOA, contact 
Customer Support.
Problem
 Cause
Action
Cradle LEDs turn red
All LEDs are flashing red
1 Disconnect cradle from 
mains power.
2 Reconnect cradle to mains 
power.
3 If problem persists, contact 
Customer Support.
Cradle orange LED is 
blinking, but cradle is not in 
discharge mode
Hardware/Software problem
PillCam recorder DR2 cannot 
be placed in cradle
Hardware malfunction
Contact Customer Support
Problem
 Cause
Action
Capsule LED blinking red
Interference
Recorder malfunction
Refer to PillCam recorder DR3 LEDs 
Indications
Capsule LED blinking white
Capsule not paired yet
SD card errors
Wrong or damaged SD 
card
Sensor connection 
Wrong or damaged 
sensor
PillCam recorder freezes
Possible static 
electricity discharge
Perform Hard Reset on the 
PillCam recorder
Contact Customer Support