Essential Performance
Capsule Labeling
Each box has a label at the bottom as shown below. Each capsule is marked with the expiration date, lot 
number, and a unique capsule ID code.
 
Essential Performance
PillCam Capsules 
ON-Mode
Data transmitting to PillCam recorder is considered to be essential performance of the PillCam capsules. 
The PillCam capsules shall transmit data continuously monitored by on-line image display as received by 
PillCam recorder.
OFF-Mode
No unintentional transmissions are allowed.
PillCam Recorder DR2 and PillCam Recorder DR3
Data receiving by PillCam recorder is considered to be essential performance of the PillCam recorder DR2 
and PillCam recorder DR3. 
Warnings
PillCam Capsule Endoscopy System and its components need special precautions regarding EMC and 
need to be installed and put into service according to the EMC information provided in the 
accompanying documents. 
Portable and mobile RF communications equipment can affect the PillCam video capsule and the 
PillCam recorder.
PillCam video capsules and PillCam recorder should not be used adjacent to or stacked with other 
equipment and that if adjacent or stacked use is necessary, the equipment or system should be observed 
to verify normal operation in the configuration in which it will be used.