Given Imaging Limited CAPDH-2 Manuale Utente

Pagina di 250
Performing PillCam Capsule Endoscopy
Real-Time Viewing with PillCam Recorder DR3 Only—Internal Mode
85
In COLON capsule procedures, the appearance of the End of Procedure icon depends on how much 
time has passed without paired capsule reception in the PillCam recorder DR3:
During the first 5 hours of the procedure, 25 minutes of no paired capsule reception.
Between 5 hours and 10 hours of the procedure, decreasing incrementally from 25 minutes to 10 
minutes of no paired capsule reception.
After 10 hours into the procedure, 10 minutes of no paired capsule reception.
At the end of the procedure, the PillCam recorder and sensors may be removed from the patient. The 
patient may then return to a normal daily routine.
Real-Time Viewing with PillCam Recorder DR3 Only—
Internal Mode
To activate real-time viewing with PillCam recorder DR3, press the middle navigation button below the 
Real-Time icon 
, then immediately press the left then right buttons one after the other.
The following screen appears on the PillCam recorder. 
֠
֠
֠
֠
Note
Using Real-time Viewing on the PillCam recorder with an external real-time 
viewer (laptop) is not supported.
1 3
2