Given Imaging Limited PATENCYSCANNER Manuale Utente

Pagina di 38
Given Imaging Ltd.
30
DOC-0062-02
4.
Ask the patient to take off heavy clothes such as jackets, coats or sweaters. 
5.
Ask the patient to stand up.
6.
Take Scanner off the hook.
7.
Verify that the Scanner is at least 5 meters away from an operating monitor 
(monitor in adjacent rooms should be turned off) or any metallic objects and 
turn it on. 
8.
Verify the following:
Power On LED is on.
Battery LED is off.
9.
While the back of the Scanner faces the patient, bring the Scanner close to the 
patient and scan his/her abdomen by moving the Scanner up and down and 
sideways. 
If capsule is not detected scan also the patient’s back and flanks by moving the 
Scanner up and down and sideways.
Note
Maximum receiving is at the handles of the Scanner. To use the 
maximum reading range, move the Scanner along the required 
scanning area including the handles of the Scanner.