  ER920W Wireless Event Monitor  
 
14 
 
6.  The Monitor is ready to record events. 
A.  The general settings of the Monitor are displayed during 
operation by the number of dots after the word 
“MONITORING” shown on the display. 
 
Display 
Setting 
MONITORING 
No Arrhythmia or AFIB 
MONITORING. 
Arrhythmia only 
MONITORING.. 
Arrhythmia and AFIB 
MONITORING 
AFIB only 
 
 
 
 
 
  ER920W Wireless Event Monitor  
 
15 
Patient Operating Instructions 
 
Display overview: 
 
 
               
 
 
 
To Connect Monitor:   
1.  Snap lead wires onto electrodes first, and then apply electrodes 
according to physician instructions.  
2.  Reapply fresh electrodes daily. 
3.  Insert the Patient Cable into the Monitor.   
NOTES: 
A.  Depending on the cable, there should be a single or double beep 
that indicates a good patient connection. If no beep is heard, 
double check cable connections.  
B.  Viewing the display just after the Patient Cable is connected 
will show the signals from each of the channels.  There will also 
be an accompanying beep for the channel number displayed. 
C.  The Monitor will not respond to any button presses other than 
Record Event during the preview time. 
D.  Square wave will be displayed on the 
LCD if a good patient connection is not 
established.  An adequate connection with 
good ECG signal must occur before the 
ER920W wireless arrhythmia monitor 
initiates monitoring.   
E.  Monitor is now looping and ready to record.