EMBLEM™ S-ICD PROGRAMMER:  DECLARATIOnS TABLES
Table 5: Declaration Electromagnetic Immunity Part 2
The Model 3200 programmer is intended for use in the electromagnetic environment specified below. The customer or the user of the Model 3200 programmer should 
assure that it is used in such an environment.
Immunity Test
IEC 60601 Test 
Level
Compliance 
Level
Electromagnetic Environment Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 V
Portable and mobile RF communications equipment should be used no closer to 
any part of the Model 3200 programmer, including cables, than the recommended 
separation distance calculated from the equation applicable to the frequency of 
the transmitter.
Recommended Separation Distance
3.
5
V
1
√ 
P
d
 
=
 
[ ]
3.
5
E
1
√ 
P
d
 
=
 
[ ]
7
E
1
√ 
P
d
 
=
 
[ ]
80 MHz to 800 MHz
800 MHz to 2.5 GHz
150 KHz to 80 MHz
where P is the maximum output power rating of the transmitter in watts (W) 
according to the transmitter manufacturer and d is the recommended separation 
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic 
site survey,
a
 should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following 
symbol:
Radiated RF
IEC 61000-4-3
3 V/m
80 Mhz to
2.5GHz
3 V/m
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, 
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess 
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the 
measured field strength in the location in which the Model 3200 programmer is used exceeds the applicable RF compliance 
level above, the Model 3200 programmer should be observed to verify normal operation. If abnormal performance is 
observed, additional measures may be necessary, such as reorienting or relocating the Model 3200 programmer.
b
 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m