CLINICAL REFERENCE TOOLS
PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft 
 
255
All information contained herein is subject to the rights and restrictions on the title page.
PatientNet Arrhythmia Analysis Software V1.05
Alarm Label
Rhythm
Description
MUSCLE
Muscle
Muscle Artifact
ASYSTOLE
Asystole
No QRS Detected for 3.0 Sec
CHKSIGNAL
Check Signal 
Intermittent no signal detected
NO SIGNAL
No Signal
No QRS Detected for 9.0 Sec
V-FIB
Ventricular Fibrillation 
Rapid disorganized Ventricular Impulses, no QRS’s
V-TACH
Ventricula r Tachycardia 
Configurable.
The number of consecutive PVCs can be set equal to, and 
between, 3 and 8. The Heart Rate can be set equal to, and 
between, 100 and 120 BPM. A V-TACH alarm is triggered 
when the consecutive PVC count is reached AND the heart rate 
is greater than or equal to the set Heart Rate Value.
V-RUN
Ventricular Run
V-RUN is triggered when the number of consecutive PVCs is 
greater than 2 and less than the V-TACH configured PVC value 
(i.e. when the V-TACH is configured at 3 PVCs, the V-RUN 
alarm is never triggered).
V-RHYTHM
Ventricular Rhythm
V-RHYTHM is triggered when the number of consecutive 
PVCs is greater than or equal to the V-TACH configured PVC 
value, but the Heart Rate is less than the V-TACH configured 
heart rate value.
BIGEMINY
Ventricular Bigeminy
N-PVC-N-PVC-N-PVC Sequence
TRIGEMINY
Ventricular Trigeminy
N-N-PVC-N-N-PVC-N-N-PVC Sequence
COUPLET
Ventricular Couplet 
2 Consecutive PVCs
HIGH PVC
High PVC
PVC Count > High PVC Limit
PVC
PVC
Single PVC
SV-TACH
Supraventricular Tachycardia 
8 Consecutive SVEs, HR 150 or More
HIGH HR
High Heart Rate 
HR Greater than High Rate Limit
LOW HR
Low Hear Rate
HR less than Low Rate Limit
NO ARR
Arrhythmia unable to analyze
No good leads are available for analysis
CHK LEAD
Check Lead
One or more of the ECG leads has a poor connection and/or is 
causing significant baseline wander
REGULAR
Normal Dominant 
Rhythm