RECORD IN USE DATE AND THE EXPIRATION DATA ON SENSOR PAD 
 
THE EXPIRATION DATA OF SENSOR PAD = PAD USABLE LIFE + IN USE DATE 
 
THE EXPIRATION DATA OF PAD TRANSMITER = 24 MONTHS + IN USE DATE 
 
4. 
Setting delay time 
There  are  optional  alarm  delay  of  0,  2,  4  seconds  on  SPT  (bed/chair  Sensor  Pad 
Transmitter only) to reduce false alarms. When used with a bed or chair sensor you 
may choose to have the alarm sound immediately when pressure is removed from 
the  sensor  or  choose  a  2  or  4  second  delay  to  accommodate  for  any  slight 
movement.   
 
Enter Setting delay time mode 
a.  Press and hold the Button (more than 2 seconds) on the SPT until you see 0s light is 
blinking.    SPT is in setting delay time mode now. 
    Time delay light (0s, 2s, 4s) 
b.  Press  and  release  the  Button  on  the  front  of  SPT,  next  Time  delay  light  will  be 
blinking.  You  can  continue  to  press  and  release  the  Button  until  desired  light  is 
blinking.   
   
Exit Setting Notification mode 
a. 
Once you’ve landed on your desired Time delay light, stop operation.   
b.  The SPT automatically exits setting delay time mode after 10 seconds of inactivity. 
This means the current Time delay is recorded to the SPT. 
5. 
Test Sensor Pad Transmitter 
a.    It  is  important  to  test  the  bed  alarm  before  using  with  a  patient.  To  test,  Apply 
pressure  to  bed/chair  Sensor  Pad,  and  released  pressure  from  pad.  The  wireless 
alarm will be transmitting to receiving devices. 
b.  If the programed receiving devices will get alarm, the Alarm light on SPT will flash to