Bionime Corporation GM232 Manuale Utente
49
48
Cleaning and Disinfecting Procedures
To clean the meter:
1. Throughly wipe the entire surface of the meter with disinfecting wipes
1. Throughly wipe the entire surface of the meter with disinfecting wipes
listed above to clean any possible dirt, dust, blood and other body
fluids.
fluids.
To disinfect the meter:
2. Take another disinfecting wipe and wipe the meter thoroughly. (Note: All
2. Take another disinfecting wipe and wipe the meter thoroughly. (Note: All
blood and body fluids should be cleaned from surface before
performing the disinfecting procedure)
performing the disinfecting procedure)
3. Allow the surface to remain wet for 1 minute.
4. Allow to air dry.
4. Allow to air dry.
NOTE
- Clean and disinfect the outside of the device only. Do not remove battery cover when
- Clean and disinfect the outside of the device only. Do not remove battery cover when
cleaning and disinfecting.
NOTE
- Your GE006 Blood Glucose Meter has been tested to ensure that there is no change in the
- Your GE006 Blood Glucose Meter has been tested to ensure that there is no change in the
performance or external materials of the device after 550 cleaning cycles and 550
disinfecting cycles. The testing simulates 2 cleaning and disinfecting cycles per week over
the typical life of the meter (5 years).
disinfecting cycles. The testing simulates 2 cleaning and disinfecting cycles per week over
the typical life of the meter (5 years).
1
Allow to remain
wet for 1 minute
wet for 1 minute
Cleaning and Disinfecting Procedures
To clean the lancing device:
1. Throughly wipe the entire surface of the lancing device with
1. Throughly wipe the entire surface of the lancing device with
disinfecting wipes listed above to clean any possible dirt, dust, blood
and other body fluids.
and other body fluids.
To disinfect the lancing device:
2. Take another disinfecting wipe and wipe the Lancing device
2. Take another disinfecting wipe and wipe the Lancing device
thoroughly. (Note: All blood and body fluids should be cleaned from
surface before performing disinfecting procedure)
surface before performing disinfecting procedure)
3. Allow the surface to remain wet for 1 minute.
NOTE
- Your
- Your
GD500 Lancing Device has been tested to ensure that there is no
change in the performance of the device after 550 cleaning cycles and 550 disinfecting
cycles. The testing simulates 2 cleaning and disinfecting cycles per week over the typical
life of the meter (5 years).
cycles. The testing simulates 2 cleaning and disinfecting cycles per week over the typical
life of the meter (5 years).
Rightest™
2
3
Allow to remain
wet for 1 minute
wet for 1 minute
1