Caution
 
5
4
Precaution
- Before using the GE006 Blood Glucose Monitoring System to test your blood glucose, please read all of 
the instructions and conduct all of the tests including the quality control test (Refer to page 37). 
- Please perform the quality control test regularly to make sure the test results are accurate.
- The GE006 Blood Glucose Meter can only be used with GE006 Blood Glucose Test Strips.  Other brands’ 
test strips should not be used under any circumstances. The use of other brands’ strips may give 
inaccurate results.
- The GE006 Blood Glucose Monitoring System is intended for in vitro diagnostic use only. The GE006 
Blood Glucose test results using fresh capillary whole blood samples from the fingertip, palm and 
forearm are calibrated to be the equivalent to that of plasma testing.
- The GE006 Blood Glucose Monitoring System is intended for self-testing. It should not be used to 
screen for or diagnose diabetes mellitus.
- If the GE006 Blood Glucose Meter and Test Strips are exposed to a substantial change in temperature, 
please wait 30 minutes before measurement.
- Follow all environmental protection regulations when disposing of batteries.
- Avoid contact with dripping or splashing liquids.
- The minimum blood sample size to test using the GE006 Blood Glucose Monitoring System is 0.75µL: 
   (     )
Blood sample size above 3.0µL might contaminate the test strip port and the meter.
Blood sample size below 0.75µL may cause an inaccurate result or may prevent a meter reading. An Er4 
reading will be displayed if the sample size is too small. In this case, repeat the test with a new test strip.
Sample Size Example
0.75µL       1.0
  1.5µL              2.0µL
µL            
             3.0µL            
Important Safety Notes:
- The GE006 Blood Glucose Meter and Lancing Device are for single patient use. Do not 
share them with anyone including other family members. 
- All parts of the kit are considered biohazardous and can potentially transmit infectious 
diseases, even after following the cleaning and disinfecting procedures. 
- Users should wash their  hands thoroughly with soap and water before and after handling 
the meter, lancing device, or test strips.
Please refer to the 
section "Cleaning and Disinfecting Procedures" on page 47.