About Quality Control Testing
About Quality Control Testing
What is a Quality Control Test ?
To ensure proper monitoring function, it is necessary to regularly perform a quality control test.
Use one of the Rightest Control Solution GC550 when testing your Rightest Blood Glucose Monitoring System  
GM232B in the Control Solution Mode. If the test result is within the Control Solution Range printed on the strip vial 
label, the
 Blood Glucose Monitoring System GM232B passes the Quality Control Test and your
 
Blood Glucose Monitoring System GM232B is working properly.
 Rightest
 Rightest
Control Solution Range:
Example of Control Solution Range printed on 
your test strip vial label.
Control Solution Range mg/dL
Level 2:              107-145 
Level 4:              315-426 
NOTE
- Level 2 of 
 is provided in your starter kit. If you want to purchase 
 or 
another 
 control solutions, please contact the GE Diabetes Customer Support Center 
.
Rightest
 Control Solution GC550
Level 4
Level 2
toll free at 
1-866-613-7085 
When should a Quality Control Test Be Performed ?
- To ensure that your
 Blood Glucose Meter 
and
 Blood Glucose Test Strip 
are   
working properly.
- To confirm that you are following the correct testing procedures.
- To prepare for your initial blood glucose test.
- To check the 
 Blood Glucose Test Strip 
when you open a new vial of strips.  
- To practice testing.
Required Items for Quality Control Tests
To perform a quality control test, prepare the items below :
Blood Glucose Meter 
 Blood Glucose Test Strips 
 Rightest
 Rightest
Rightest
To check your Rightest Blood Glucose Meter 
 after it has been dropped, damaged or exposed to liquids.
- If you suspect that your test results are inaccurate, or if your test results are not consistent with the way you feel.
Rightest 
Rightest
Rightest 
Control Solution GC550
GM232B 
GM232B 
GM232B 
GM232B
GM232B
GM232B