Recalling Average Test Results
Caring for Your Meter 
5. Quick Searching:
    To view test results in sequential order automatically, enter the Memory Mode. Press the Right or Left button for 2 
seconds. (The Right button is for viewing from the most current to the oldest test results. The Left button is for viewing 
from the oldest to the most current test results). Release the button to stop at any particular test result.
NOTE
 Test results must exist during the desired 
time interval. For example: To get a 14-day average on 1/30/14, you must have test results dated between 1/17 
and 1/30. If no test results exist during that time frame you will not receive an average.
- To exclude test results from average calculation, you need to remain in Test Mode and perform the exclusion 
right after you get the test result. Once you move to MEM Mode, Average Mode or Time Mode, you will not be 
able to perform the exclusion.
- The "Lo", "Hi" results, control solution test results and test results under abnormal temperature < 50°F (10°C), 
> 104°F (40°C) will be excluded from average calculations.
The Average function requires that the correct time and date is set.
Maintenance
Keep your meter and test strip free of dust, water or any other liquid. Store the meter in the carrying case when not 
in use. If your meter is dropped or damaged, perform a quality control test with the control solution before 
performing a blood glucose test.
Cleaning Meter
Clean the outside of the meter with disinfecting wipes listed on next page. Do not get the test strip port wet.
Cleaning Test Strip Port
If your test strip port is stained with blood, control solution or any liquid, please use disinfecting wipes listed on next 
page to clean it immediately. Do not use anything wet to clean it. Perform a quality control test to ensure the