Performing a Blood Glucose Test
Performing a Blood Glucose Test
16. Pull off the depth adjustable cap. Without touching the 
used disposable lancet, insert the lancet tip into the 
protective lancet cover.
17. Hold the release button in one hand (see picture 17, step 1) 
and pull on the plunger with the other hand (see picture 
17, step 2) to safely eject the used disposable lancet.
19. Replace the depth adjustable cap after finishing the test.
18. Discard the used disposable lancet into an appropriate 
puncture-proof or biohazard container.
18
16
17
19
2
1
CAUTION
- Do not apply your blood drop to the sample port on the strip until you see the "      " appear. The Rightest 
Blood Glucose Meter 
is performing an internal test and will display "      " and "           " if you apply 
blood too soon. If this occurs, please repeat the test with a new test strip.
- Record the opening date of a new test strip vial. Discard the vial of test strips 3 months after opening.
- Always keep the metal contacts of the test strip port clean. If any dust or impurities are 
present, please clean with a small, soft brush.
- The
 
Blood Glucose Monitoring System GM232C and Rightest GD500 Lancing Device are intended 
for a  single user only. Do not share these devices with anyone, including other family members.
- All parts of this kit are considered biohazards and can potentially transmit infectious diseases, 
even after you have performed the cleaning and disinfecting procedure.
- Users should wash hands thoroughly with soap and water after handling the meter, lancing 
   device, and test strips.
- Please refer to the section "Cleaning and Disinfecting Procedures"on page 48 for cleaning and 
  disinfecting instructions.
GM232C 
 Rightest